Method
Discrete method, single line
Composition
Analyzer, (sampler, data processor)
Racks for the data processor (PC, monitor, keyboard, pointing device) shall be fixed on the sampler and the monitor shall be installed on the arm.
Sample Types
Serum, urine, whole blood (Only for HbA1c analysis), plasma
Sample availability is defined by the sample viscosity. Whole blood can be used with sampler option.
Item Numbers for Simultaneous Analysis
A Maximum of 60 items (A maximum of 63 items with ISE option)
Maximum Reagent Bottle Numbers First reagent: 60 bottles Second reagent: 48 bottles (Using 60, 30, 15 mL bottles)
Throughput
A maximum of 800 tests/hour (A maximum of 1200 tests/hour with ISE option)
800 samples/hour can be processed when performing 1item/sample continuous order.
100 tests/hour can be processed when performing HbA1c analysis only.
Analysis Method
1. Photometric method
End-point method
Rate method
Fixed method
2. Electrode method (ISE option)
Reaction Liquid Volume
120 L ~ 425 L (5 6 mm)
Reaction Time
When 3-part reagents are used, R1 is divided into R11 and R12. R1 and R11 will be discharged during the same timing interval.
Reaction Temperature
37 C ( 0.3 C)
Ver.7
Items can be Analyzed
1. Protein, colloid reaction
Total protein, albumin, TTT, ZTT
2. Enzymes and other related substances
CK, CK-MB, AST, ALT, LDH, ALP, GTP, MAO, CHE, guanase, ADA, LAP, AMY, lipase, ACP, TR-ACP, HBD, NAG
3. Low molecular nitrogen compound
Creatine, creatinine, UA, UN
4. Glucide and other related substances
Glucose, fructosamine, sialic acid, (Hb, HbA1c with sampler option)
5. Lipid and other related substances
TG, phospholipids, NEFA, cholesterol, free cholesterol, HDL-C, total bile acid, apolipoprotein, -lipoprotein
6. Electrolytes
Mg, Ca, IP (Na, K, Cl with ISE option)
7. Biological trace metals
Fe, TIBC, UIBC, Cu
8. Biological dye-related substances
Total bilirubin, direct bilirubin, indirect bilirubin
9. Drugs
Phenobaribital, phenytoin, carbamazepine, valproic acid, primidone, teophylline
10. Immunoglobuline
IgG, IgA, IgM
11. Complements and other related substances
C3, C4
12. Plasma protein
CRP, ferritin, Tf
13. Tumor-related antigens
CEA, AFP
14. Serum information
Icterus, hemolysis, lipemia
The items listed above are examples of items that can be measured.
Serum Information
Graded index indications are possible by measuring the degree of icterus, hemolysis, and lipemia
The judgment levels are user defined.
The degree of incidence can be displayed and saved to the analyzed data as a flag.
Reception of Analysis Requests
1. Combination of individual input and batch input is possible for item selection.
2. Input operation Touchscreen Pointing device Keyboard Online
3. Maximum number of requested samples
9999 samples/index
4. Maximum of 63 items can be ordered per specimen, including ISE option (when installed). When a separate reagent for LIH is used, the maximum number of ordered tests per sample is inclusive of LIH.
5. Patient information can be entered separately from item selection.
Sample Search
1. It is possible to check for the sample status by sample ID and sample No. for received samples.
2. Displays the estimated time to results output for a sample in process.
Repeat Run
1. Automatic repeat run
2. Manual repeat run is possible.
3. It is possible to change the sample volume.
Normal repeat run, condensation repeat run, dilution repeat run (including pre-dilution repeat run)
4. Reflex Testing
Rapid repeat run is possible for samples judged as repeat run with automatic repeat run. Automatic repeat run is not possible when connected to a lab automation system.
AU-C indicates and extracts samples for repeat run specimens when connected. A host computer instructs sample extraction and repeat run when connected. Only normal repeat run is available for HbA1c (whole blood) analysis.
Reflex testing function determines additional testing items from the result of one analysis and performs further analysis.
Warm Up and Standby
1. The warm-up time is 20 minutes.
2. Automatic power ON/OFF is available.
The automatic power ON/OFF function does not control the water tap.
Input Method
1. Touch screen
2. Pointing device
3. Keyboard
4. Online
Monitor Display
Japan: Japanese
Other than Japan: 13 languages will be applied
Output Method
1. Monitor display
2. Hard disk
3. Online
4. External storage device (Option)
5. Printout (Option)
STAT Operation (STAT table)
1. Starting analysis by simple operation is possible.
2. Interruption is possible between tests.
Automatic repeat run function for samples with barcodes is possible.
Communication System
Communication with HIS/LIS
The following functions are prospective options (Note 1):
1. PROService via an internet connection
For accounting and remote service systems
Sample Pre-Dilution Analysis
1. Sample dilution factor: 3, 5, 10, 15, 20, 25, 50, 75, 100 fixed
2. Diluent: User setting is possible (Only one type)
3. This can be performed during normal analysis.
Safety and Crash Detection
1. Crash detection function is available for sample and reagent transfer mechanisms.
2. Function to prevent users from touching the probe end accidentally (setting up walls etc.) is available.
Reliability
Sample probe clot detection function
Machines we support for service, maintenance, repair and installation
Beckman coulter chemistry analyzer: AU 5800, AU 680, AU 480, DXC 800, DXC 600,LX20
OLYMPUS chemistry analyzer : AU 5400, AU 2700, AU 640, AU 400
Abbott Elisa analyzer: ARCHITECT i1000 SR, ARCHITECT i2000 SR
Beckman Coulter Access 2
Horiba Pentra C400 Biochemistry
Dynex Technologies : DSX Automated ELISA System
ELISA Instruments : Dynex Reader, Washer. Seac Alisei Automatic Elisa Analyser, Seac Brio semi Automatic Elisa Analyser, Biotec Elx 800 Reader, Elx 50 Washer, Grifols Triturus Automatic Elisa Analyser, Euroimmun ANA 1 Elisa analyser, Orgentec Alegria elisa analyser.
Urine Instruments: Iq 200 Urine Analyser ( Arkray and Velocity models )