1. Method
● Specifications
Discrete, Single-line • multi-test, Patient-oriented
2. Construction
● Specifications
Analyzer • data processor • printer
● Remarks
The data processor is a personal computer.
3. Sample type
● Specifications
Serum • urine • plasma
● Remarks
For each item, the analysis parameters can be set as serum, urine, or one other type.
4. Number of simultaneous analysis items
● Specifications
Item selection input : Max. 51 items (including 3 ISE items)
At the time of special reagent use, LIH is included in the number of items, but it is not included when not used.
Number of reagent bottles which can be set : Max. 38/all 2 steps
Max. 76/all 1 step
● Remarks
99 analysis condition items are memorized., 2 items can be set per bottle.
5. Processing capacity
● Specifications
Main unit : 400 tests/h
When Na, K, Cl/ISE 3 items are connected : Max. 800 tests/h
● Remarks
ISE alone : 200 samples/h
6. Analysis method
● Specifications
(1) One point method-end point
(2) Two point method-automatic blank compensation at the end point
(3) Rate method
(4) Fixed method
(5) Electrode method (ISE)
● Remarks
ISE is optional.
7. Reaction fluid volume
● Specifications
150 to 550 L
8. Reaction time
● Specifications
Max. 8 min, 37.5 sec. (2 steps)
9. Reaction temperature
● Specifications
37C
● Remarks
Sample temperature accuracy 0.3 C
10. Items which can be analyzed
● Specifications
(1) Colorimetry
Enzymes/lipids/proteins/sugar/nitrogen compounds/inorganic substances/complements/others (TBA etc.)
(2) Turbidimetry
lgG/lgA/lgM/C3/C4/RF/CRP/ASO/transferrin/others
(3) Latex agglutination
RF/CRP/ASO/others
(4) Homogeneous EIA (EMIT)
DAU/TDM/others
(5) ISE (option)
Na/K/Cl
● Remarks
ISE is optional.
11. Serum information
● Specifications
The degree of chyle, jaundice, and hemolysis can be measured, and staged indication is possible.
A method using exclusive fluid or a method using a fluid showing no absorption in the visible range can be
selected.
● Remarks
The judgment level for stepped indication can be set.
12. Reception
● Specifications
(1) Individual and group item selection can be specified together.
(2) Item selection input method
1) Keyboard
2) On-line
3) Mouse
4) Touch screen (option)
(3) For item selection, input is possible for max.
8000 samples/index.
Max. 90 indices can be stored.
(4) Max. 51 items can be received per sample.
(5) Input of patient information is possible separate from item selection.
● Remarks
A total of 30,000 samples can be memorized.
13. Reanalysis
● Specifications
(1) Automatic reanalysis and manual reanalysis are possible.
(2) Sample quantity : Normal, increased, decreased
14. Warmup & Standby
● Specifications
(1) Use with standby at all times is possible.
(2) Warmup time : 20 min.
(3) Power supply automatic start-up/shut-down possible (fixed sequence)
● Remarks
At the time of power failure, the automatic start-up function does not recover.
15. Input method
● Specifications
(1) Key input interactive with the CRT
(2) On-line
(3) Mouse
(4) Touch screen
● Remarks
CRT display
Japan - Japanese
USA - English
Europe - English
General areas - English
16. Output method
● Specifications
(1) CRT display
(2) Printer printing
(3) Hard disk
(4) Floppy disk
(5) On-line (RS232C)
17. STAT operation
● Specifications
(1) Start by one-key operation
(2) Interrupt between tests
● Remarks
Under some conditions, additional key operation is required.
18. Modem
● Specifications
Communication of various information by telephone lines is possible (option).
19. Predilution
● Specifications
(1) Dilution factor : 5 to 100
(2) Diluent : Water
(3) Execution possible in normal analysis mode
(4) Object samples : Serum, Urine, one other type
20. Safety
● Specifications
(1) With collision prevention function for sample and reagent transfer devices
21. Reliability
● Specifications
(1) Sample dispensing clogging detection
Machines we support for service, maintenance, repair and installation
Beckman coulter chemistry analyzer: AU 5800, AU 680, AU 480, DXC 800, DXC 600,LX20
OLYMPUS chemistry analyzer : AU 5400, AU 2700, AU 640, AU 400
Abbott Elisa analyzer: ARCHITECT i1000 SR, ARCHITECT i2000 SR
Beckman Coulter Access 2
Horiba Pentra C400 Biochemistry
Dynex Technologies : DSX Automated ELISA System
ELISA Instruments : Dynex Reader, Washer. Seac Alisei Automatic Elisa Analyser, Seac Brio semi Automatic Elisa Analyser, Biotec Elx 800 Reader, Elx 50 Washer, Grifols Triturus Automatic Elisa Analyser, Euroimmun ANA 1 Elisa analyser, Orgentec Alegria elisa analyser.
Urine Instruments: Iq 200 Urine Analyser ( Arkray and Velocity models )